1. After every patient treatment, wipe and rinse the handpiece with cold tap water.
  2. Remove the handpiece from the cord.
  3. Prior to reprocessing, clean the water channel of your handpiece by blowing it out with air.
  4. Prior to reprocessing, clean the air-powder channel of your handpiece:
    1. Fill a syringe with water.
    2. Connect the EasyClean and the handpiece.
    3. Rinse the air-powder channel.
    4. Repeat if neccessary.
  5. Follow the Reprocessing Instructions to clean the instrument.
  6. Package the cleaned and disinfected handpiece in a single or double pouch.
  7. Sterilise immediately.