1. After every patient treatment, remove the PERIOFLOW® Nozzle using the Check Tool.
    ! DO NOT remove the Nozzle by hand.
  2. Wipe and rinse the handpiece with cold tap water.
  3. Remove the handpiece from the cord.
  4. Prior to reprocessing, clean the air-powder channel of your handpiece:
    1. Fill a syringe with water.
    2. Connect the EasyClean and the handpiece.
    3. Rinse the air-powder channel.
    4. Repeat if neccessary.
  5. Follow the Reprocessing Instructions to clean the instrument.
  6. Package the cleaned and disinfected handpiece in a single or double pouch.
  7. Sterilise immediately.