The Check Tool helps you check your instrument's wear and remove the PERIOFLOW® Nozzle. The following instruments can be checked:
- P/PS Instrument
- PI Max Instrument
- PSL/PSR Instrument
- Remove the sticker.
- Place the tip of the instrument horizontally into the dedicated opening.
- If the tip of the instrument reaches the inside of the circle, you can continue using the instrument. If not, you will need to replace it.