Every Morning:

  1. Place your filled water bottle onto the device.
  2. Remove the handpiece.
  3. Set the water to 10 on the black dial.
  4. Turn the device on.
  5. Remove the cord and place it over a sink or bowl.
  6. Press the foot pedal for 30 seconds.
  7. Water will start coming out of the cord. It will stop, once you stop pressing the foot pedal.
  8. Your AIRFLOW® device is now ready for use.

Between Patients:

  1. Remove the handpiece and sterilise it.
  2. Clean the outside surface of your device with a cloth and disinfectant.
  3. Repeat this process after every patient.

Every Evening:

  1. Fill the blue bottle with Night Cleaner and place it onto the device.
  2. Remove the handpiece.
  3. Set the water to 10 on the black dial.
  4. Remove the cord and place it over a sink or bowl.
  5. Press the foot pedal for 30 seconds.
  6. The Night Cleaner solution will start coming out of the cord. It will stop, once you stop pressing the foot pedal.
  7. Your AIRFLOW® device is now clean.