1. Remove the handpiece and press the foot pedal.
  2. If air is coming out, it means your handpiece is clogged.
  3. After every use, it is important to flush out your AIRFLOW® and PERIOFLOW® lumens.
  4. Use a syringe filled with water, attach it to the EMS EasyClean and insert into your handpiece.
  5. Empty the syringe to flush your handpiece.
  6. Repeat this process 2-3 times.
  7. If your handpiece is still clogged, soak it in warm water for 20 minutes and then use the EasyClean to flush it again.
  8. The next thing to check, is your Powder Chamber.
  9. Release the Powder Chamber from the device.
  10. Place your finger over the hole of your Powder Chamber and shake it to loosen up the powder.
  11. Check if this resolves the issue.
  12. If not, you need to remove the cap from the Powder Chamber and remove all powder.
  13. Blow compressed air into the Powder Chamber to clean it out.
  14. If this also does not fix the issue, and you see that powder is being released from the cap after pressurisation, or the chamber lighting up and blinking with a white colour, the problem might be with the cord.