- Fill a glass with the cleaning solution (liquid acetic acid 20% + 2-3 drops of dishwashing detergent or soap).
- Soak the AIRFLOW® MAX handpiece in the solution.
- Put the glass with the handpiece into the water-filled ultrasonic bath.
- If possible, set the temperature to 50°C (122°F) and activate the ultrasonic power at your usual setting for approximately 15 minutes.
- Rinse the AIRFLOW® MAX handpiece with tap water.
- Place the handpiece in a glass half-filled with distilled water.
- Repeat the ultrasonic process for approximately 3 minutes.
- Afterwards, start the Reprocessing.
If you happen to have more issues, please contact your EMS Service Centre.