1. Fill a glass with the cleaning solution (liquid acetic acid 20% + 2-3 drops of dishwashing detergent or soap).
  2. Soak the AIRFLOW® MAX handpiece in the solution.
  3. Put the glass with the handpiece into the water-filled ultrasonic bath.
  4. If possible, set the temperature to 50°C (122°F) and activate the ultrasonic power at your usual setting for approximately 15 minutes.
  5. Rinse the AIRFLOW® MAX handpiece with tap water.
  6. Place the handpiece in a glass half-filled with distilled water.
  7. Repeat the ultrasonic process for approximately 3 minutes.
  8. Afterwards, start the Reprocessing.

If you happen to have more issues, please contact your EMS Service Centre.