How to check the wear of the PI MAX instrument?

  1. Insert the instrument tip into the opening of the tool that is marked with 'check'.
  2. If the instrument tip is not visible in the side opening of the tool, it needs to be replaced.

How to disconnect the PI MAX instrument tip?

  1. Insert the instrument into the opening of the tool marked with 'disconnect'.
  2. Insert the screw until it is fully connected to the tool.
  3. Remove the black tip of the instrument.
  4. Remove the screw.
  5. You can now remove the instrument.

How to connect a new PI MAX instrument tip?

  1. Insert the black tip through the small whole of the instrument
  2. Insert the instrument into the opening of the tool marked with 'connect'
  3. Insert the screw until it is fully connected to the tool.
  4. Remove the screw.
  5. You can now remove the instrument.